Programming the Neurostimulator
1
2
To adjust cycling and ramping:
Cycling and ramping settings are active when settings have been
set to the Neurostimulator.
Note: The cycling and ramping settings are not active during test
stimulation. To test cycling and ramping, save the stimulation settings to the
Neurostimulator by pressing "Set Device Therapy" then activate stimulation
on the Patient Device screen.
Press the button for the parameter and a pop-up will appear for
adjusting the setting
Cycling
o Allows stimulation to turn on and off automatically at
specified intervals.
o The time that stimulation is on and off are independently
programmed.
o On and off time can be programmed from 0 (off) to 59
seconds, 1 minute to 59 minutes, and 1 hour to 24 hours.
Note: Cycling time can only be programmed to values longer than
Ramping time.
Ramping
o Allows stimulation to ramp up and down between zero
amplitude and the targeted stimulation amplitude when
stimulation turns on and off.
o Ramping time can be programmed at "Off", 1, 2, 4, 8, 15,
or 30 seconds.