Resetting the Trial Stimulator
The Stimulator can be reset if there is a need to not deliver
stimulation for a period of time (e.g., days or longer) or if it was
configured with the wrong trial type. Resetting the EPG puts the
Stimulator in a power‐ conserving state to preserve the battery
of the Stimulator and erases all patient data from the Stimulator.
How to reset the Stimulator: Press the "Reset EPG" button ( 1 ).
A prompt will ask for confirmation before resetting the
Stimulator. The Stimulator will automatically disconnect from
the CP and the CP will return to the Home screen when the
Stimulator is reset.
1
Note: The CP and Patient Remote Control will not be able to communicate
with the Stimulator until the Stimulator is activated by pressing the button
on the back of the Stimulator and the Remote Control serial number is
programmed by the CP.
How to wake up the Stimulator: Press the button on the back
of the Stimulator. The green light will flash next to the button.
Connect with the CP within 90 seconds or the Stimulator will
turn off.