Herunterladen Diese Seite drucken

Axonics 2501 Handbuch Seite 54

Programmiergerät
Vorschau ausblenden Andere Handbücher für 2501:

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 92
Programming the Trial Stimulator
Electrode Configuration
1
The CP provides the ability to use automatically generated
recommended electrode configurations or to manually set which
2
electrodes are active and inactive during stimulation.
Electrode Recommendations
The Programming screen displays up to four recommended
electrode configurations ( 2 ). These recommendations are
intended to provide directional guidance regarding which
electrode configurations to test. For tined lead stimulation,
the recommendations are generated based on the stimulation
thresholds and motor and sensory responses recorded during
tined lead placement. Ultimately, electrode configuration
selection should be based on patient comfort and symptom
reduction. To set the electrode configuration to a recommended
configuration, press the image of the recommendation.
Note: PNE lead electrode recommendations reflect all available electrode
configurations.
Note: If tined lead thresholds are not saved to the Stimulator, default
electrode recommendations are provided.
50

Werbung

loading

Verwandte Produkte für Axonics 2501

Diese Anleitung auch für:

1601190112011101