Programming the Neurostimulator
68
1
6
3
2
5
4
Frequency: the stimulation frequency can be adjusted from
2‐130 Hz. Adjustments are made in 1 Hz increments from 2‐50
Hz and 5 Hz increments from 50‐130 Hz.
Amplitude Step Size
The baseline stimulation level is set by the stimulation level
slider. When therapy is saved to the Neurostimulator, the
stimulation amplitude shown will be the default stimulation
amplitude.
The amplitude step size sets the increment by which the
stimulation level increases or decreases when up or down is
pressed on the Remote Control. Use the arrows to the left and
right of the Amplitude Step Size value to adjust it lower or higher.
Note: The amplitude step size must be set when programming the
Neurostimulator. The Remote Control allows for 3 steps to increase from
the baseline stimulation amplitude to the maximum amplitude. Test the
stimulation range from the desired amplitude to the maximum amplitude
to ensure it is comfortable for the patient.
Pulse Width can be set between 60 and 450 μs, adjustable in
10 μs increments.