Programming the Neurostimulator
Delivering test stimulation
4
To deliver test stimulation, set the desired stimulation amplitude
on the stimulation bar ( 1 ) (see manual section Controlling
Stimulation Amplitude for more information). Then press the
stimulation button to deliver stimulation ( 2 ). The stimulation
button will light up when Stimulation is on.
1
When stimulation is on, the electrode configuration, cycling
setting, ramp setting, frequency, and pulse width cannot be
changed. Incremental changes can be made to the stimulation
amplitude.
3
Save Therapy
2
After therapy settings have been configured, press "Set Device
Therapy" ( 3 ) to save the settings to the Neurostimulator.
To exit, press "Patient Device" to return to the Patient Device
screen ( 4 ).
Note: If "Set Device Therapy" is not pressed prior to exiting, the
Neurostimulator settings will revert to last saved settings when the
Programming screen is exited. When the Programming screen is exited,
a prompt will appear requesting confirmation to exit without saving the
therapy settings.
70