TEDS transducer identification

8
TEDS transducer identification
TEDS (Transducer Electronic Data Sheet) allows you to store the sensor
characteristic values in a chip as per IEEE1451.4. The C10 can be delivered
with TEDS, which is then fitted in the transducer housing, connected and
supplied with data by HBM before delivery. If the force transducer is ordered
with TEDS, the characteristics values from the manufacturing certificate are
stored in the TEDS chip, and if an additional DKD calibration is ordered, the
calibration results are stored in the TEDS chip.
The TEDS module is connected between PIN E (sense lead (­)) and PIN D
(excitation voltage lead (­)). HBM's zero wire configuration allows the TEDS to
be read without an additional sensor cable.
If a suitable amplifier is connected (e.g. QuantumX from HBM), then the
amplifier electronics will read the TEDS chip and parameterization will follow
automatically, without any intervention required by the user.
The chip content can be edited and configured with suitable hardware and
software. This can be implemented. with the Quantum Assistant, for instance,
or with catman DAQ software from HBM. Please comply with the operating
instructions of these products.