Herunterladen Inhalt Inhalt Diese Seite drucken

HBM C10 Montageanleitung Seite 34

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Versions and ordering numbers
Number of
measuring bridges
Sensitivity
Calibration
Transducer
identification
Mechanical design
Plug protection
Electrical
connection
bridge A
32
For reasons of redundancy, it is necessary in devices
relevant to safety to check the plausibility of the
measurement signal with a second measuring bridge
electrically isolated from the first one on the same
measuring body. This makes it possible to connect two
amplifiers working independently of one another.
The exact sensitivity is always stated on the type plate
and on the manufacturing certificate. The C10 can be
adjusted to a sensitivity of 2 mV/V (nominal (rated) forces
2.5 kN to 10 kN), or 4 mV/V (all other nominal (rated)
forces)
The sensitivity of the standard version of the C10 is more
than 4 mV/V for nominal (rated) forces from 25 kN.
(>2 mV/V for nominal (rated) forces 2.5 kN to 10 kN). If
required, you have the option to calibrate the transducers
to half the nominal (rated) force, so that the output signal
for the calibration force is also halved.
TEDS integration (integrated data sheet storing the
characteristic values of the sensor) as per IEEE1451.4,
see Chapter 8 TEDS transducer identification, Page 27.
The C10 is delivered with an adapter as standard. Upon
request, we can deliver the sensor without the foot
adapter to reduce the construction height. The require­
ments relating to the surface quality (flatness, hardness)
of the construction element on which the C10 is mounted
are thus increased.
Mechanical protection provided by fitting an additional
square section around the connector. External
dimensions (WxHxD) in mm: 30 x 30 x 20.
The standard version is a bayonet connector
(PT02E10­6P-compatible). The option is also available to
fit a screw-type device plug (PC02E10­6P­compatible). A
third variant where the force transducers are fitted with a
fixed cable is also available. In this version, all C10s with
a nominal (rated) force greater than/equal to 25 kN
achieve protection class IP68.
A04036_01_YI0_03 HBM: public
C10

Werbung