The WR Table
The word recognition (WR) table allows f or measuring multiple WR scores using dif f erent parameters (e.g.
Transducer, Test Type, Intensity, Masking, and Aided).
Upon changing Transducer, Masking, and/or Aided re-testing an additional WR entry will appear in the WR
table. This allows f or multiple WR measurements to be shown in the WR table.
Please ref er to the Callisto Additional Inf ormation document f or more inf ormation about Word Recognition
testing.
Binaural and Aided options
To perf orm binaural speech tests:
1. Click on either SRT or WR, to choose the test to be conducted binaurally
2. Ensure that the transducers are set up f or binaural testing. For example, insert Right in channel 1
and insert Lef t in channel 2
3. Click on
4. Proceed with the test; when stored, results will be stored as binaural results
To perf orm an aided test:
1. Select the desired transducer. Typically, aided testing is done in the Free Field.
However, in certain conditions, it could be possible to test deeply inserted CIC
hearing instruments under headphones, which would show ear-specif ic results
2. Click on the Aided button
3. Click on the Binaural button if the test is done in the Free Field so that the results
are stored f or both ears at the same time
4. Proceed with test; results will then be stored as aided by showing an Aided icon