Operation

Connect interfaces

The USB interface cable supplied is 1 m long. We advise against
using longer cables.
A longer cable might result in data transmission errors.
The device is switched off.
Insert the plug on the interface cable into the interface on the back of the stirrer.
Insert the other end into the USB port on your PC.
Warning: Automatic start! Risk of injury! Risk of entangle-
ment!
Rotating items can lead to severe injuries by drawing in long hair,
loose clothing, jewelry or the like.

Link interfaces

The PC is switched on.
Switch the overhead stirrer ON.
Start rotation on the PC.
The data link between the PC and stirrer is completed.
The stirrer is switched to remote control mode.
All commands are executed via the PC.
Only the slider on the stirrer is still active to allow you to end the stirring process.
You can always switch it OFF using the ON/OFF switch on the stirrer.

Interface commands

Command to
Feedback
stirrer*
r\r\n
RPM: xxxx\r\n
s\r\n
SET: xxxx\r\n
m\r\n
NCM: xxxx\r\n
FLT: No Error!\r\n
FLT: Motor Error!\r\n
f\r\n
FLT: Motor Temperature!\r\n
FLT: Stopped Manually!\r\n
FLT: overload!\r\n
Keep long hair, jewelery or loose clothing away from rotating
parts.
Before you activate remote control, place a warning sign close
to the stirrer. Every user must be aware of the potential risk of
injury.
Keep a safe distance from the device to avoid contact with any
rotating parts.
Feature
Display actual speed (rpm)
Display set speed (rpm)
Display torque (Nmm)
No error detected
Motor error
Motor overheated, restart rotation
Stirrer stopped manually
Motor overloaded