Operation
Command to
Feedback
stirrer*
HT:100P\r\n
T\r\n
HT:200P\r\n
HT:400P\r\n
Rxxxx\r\n
Rxxxx\r\nSET:xxxx\r\n
R0000\r\n
R0000\r\nSet\:0\r\n
A direction\r\n
A\r\n
B direction\r\n
A direction\r\n
B\r\n
B direction\r\n
C\r\n
Clear Error\r\n
D\r\n
Direct control\r\n
N\r\n
Null reference\r\n
F\r\n
F\r\n
M\r\n
M\r\n
t\r\n
TMP: xxx\r\n
zSET: xxxx\r\nzACT:
z\r\n
hhmmss\r\n
*RS 232 interface parameters: 9600 Baud, No Parity, 8 Bit, 1 Stopbit
Do not send commands as package,
minimum pause 0.1 seconds
Commands M, F, T and z are not valid for stirrer RZR
xxxx stands for a 1 - 4 digit figure
Receiving \r\n may be displayed in a row
Feature
Connected to Hei-TORQUE 100
Ultimate
Connected to Hei-TORQUE 200
Ultimate
Connected to Hei-TORQUE 400
Ultimate
Start rotation at set speed xxxx (rpm)
Stop stirrer, set speed saved
Switch to gear II
Switch to gear I
Delete "Overload" error message
Deactivate communication link:
Remote is control deactivated
Device can be operated
manually
Set actual torque to zero:
Torque is being calibrated
Check communication link
(Stirrer will shut down off after 60
seconds if no commands are received)
Identify device:
Display on the linked stirrer
will start flashing
Display temperature
Timer: display set time and remaining
time