Operation

Interface


Only Hei-TORQUE Precision
You can connect the stirrer to a PC via the interface. Two serial interfaces are located on
the back of the device. (Operation: see separate manual)
The following interfaces are available:
USB interface: Micro USB type AB
RS232 interface, SUB D9
You can connect the device either via USB
or via the RS232 interface.
Up to 4 overhead stirrers can be driven via
numerous PC interfaces.
➜ Close the dust protection cap on the
interface if not in use.
Software
Installation and operation of the software program is described in a separate manual.
Utilize USB-Port according to USB 2.0 Spec. up to 500 mA maximum.

Connecting and linking interfaces

Warnung:
Danger of electric shock!
If the interface inputs and outputs are supplied with excessively high
voltage and are not sufficiently insulated, in the event of a fault,
metallic parts (e.g. the housing), will be electrified.
Voltage inputs and outputs above 25 V AC or 60 V DC are deemed safe
if separated according to DIN EN 61140, or by double or reinforced
insulation in accordance with DIN EN 60730-1 or DIN 60950-1.
Only use shielded connection cables. Connect shield to connector case.
Caution:
Possible damage of device
If device sockets are not covered, fluids can get in contact with
Interface connectors. Penetrating fluids can lead to malfunctions of
interface connection and could initiate a short circuit.
Sockets at the backside of device must be covered at all times.
USB interface dust
protection cap
RS232 interface
USB interface
36