Herunterladen Inhalt Inhalt Diese Seite drucken

Interfaces; Connect And Link Interfaces - Heidolph Hei-TORQUE Expert Betriebsanleitung

Vorschau ausblenden Andere Handbücher für Hei-TORQUE Expert:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Operation

Interfaces

Hei-TORQUE Ultimate only
You can connect the stirrer to a PC via the interface. Two serial interfaces are located on
the back of the device. (Operation: see separate manual).
The following interfaces are available:
USB interface: Micro USB type AB
RS232 interface, SUB D9
You can connect the device either via
USB or via the RS232 interface.
Up to 4 overhead stirrers can be
driven via numerous PC interfaces.
Close the dust protection cap on
the interface if not in use.
Software
Installation and operation of the software program is described in a separate manual.
Utilize USB port according to USB 2.0 Spec. up to 500 mA
maximum.

Connect and link interfaces

Warning: Danger of electric shock!
If the interface inputs and outputs are supplied with excessively high
voltage and are not sufficiently insulated, in the event of a fault,
metallic parts (e.g. the housing), will be electrified.
Voltage inputs and outputs above 25 V AC or 60 V DC are deemed
safe if separated according to DIN EN 61140, or by double or
reinforced insulation in accordance with DIN EN 60730-1 or DIN
60950-1.
Only use shielded connection cables. Connect shield to connector
case.
Caution: Possible damage of device
If device sockets are not covered, fluids can get in contact with
Interface connectors. Penetrating fluids can lead to malfunctions of
interface connection and could initiate a short circuit.
Sockets at the backside of device must be covered at all times.
USB interface dust
protection cap
RS 232 interface
USB port
93