XTP601 Bedienungsanleitung
Register Type I - Status / Fehler
15
14
13
12
r
r
r
Bit
HEX
0
0001
1
0002
2
0004
3
0008
4
0010
5
0020
6
0040
7
0080
8
0100
9
0200
10
0400
11
0800
12
1000
13
2000
14
4000
15
8000
97313 DE Ausgabe 7, März 2020
11
10
9
r
r
r
r
Displays O2HSR or O2 depending on setting
%O2 out of range (beyond calibration
range, e.g. 0-25%)
Ext Comp i/p signal error
(input < 3.6mA or > 21mA)
Ext sens. signal error
(input < 3.6mA or > 21mA)
(not within ± 0.5 degC of setpoint for
continuous period of 5 minutes)
Cell T sensor error
(cell tempr measures <-50 or >80°C)
Press sensor error
(Press sensor < 850 or >1100 mbar)
(Vcomp <=1 or >=8191)
PCB tempr too high
(PCB tempr > Cell tempr setpoint)
Instrument is BLIND version (system)
8
7
6
5
r
r
r
Meaning
(system)
AL1 ON
AL2 ON
Cell T not stable
O2 sensor error
N/A
N/A
N/A
N/A
4
3
2
r
r
r
r
Namur LED
N/A
N/A
YELLOW 1 ON
YELLOW 2 ON
RED FLASH
(priority2)
RED FLASH
(priority2)
RED ON
(priority1)
RED ON
(priority1)
RED ON
(priority1)
RED ON
(priority1)
RED ON
(priority1)
N/A
N/A
N/A
N/A
N/A
1
0
r
r
50