Remove product from service before you perform the preventive maintenance inspection. Check all items listed during
EN
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance
checks more often based on your level of product usage. Service only by qualified personnel.
Note: Clean and disinfect the exterior of the support surface before inspection, if applicable.
Inspect the following items:
All fasteners are secure
Siderails move and latch properly
Casters lock with brake pedal applied
Casters are secure and swivel properly
Casters are free of wax and debris
Brake mechanism operates as intended
Steer function working as intended
Check for cracks
Fowler raises, lowers, and latches securely
Gatch raises, lowers, and secures (optional equipment)
Trendelenburg/Reverse Trendelenburg operates as intended from all locations
Ground chain intact
No leaks at hydraulic connections
Hydraulic jacks are holding
Body restraints work as intended (optional equipment)
IV pole is intact and adjusts and latches in all positions (optional equipment)
Oxygen bottle holder is intact and operates as intended (optional equipment)
No rips or cracks in the cover
Accessories and mounting hardware are in good condition and working as intended
Battery powered functionality (optional equipment)
Cables are not worn or pinched (optional equipment)
Power cord and plug are free of damage (optional equipment)
All electrical connections are tight (optional equipment)
All grounds secure to the frame (optional equipment)
Ground impedance not more than 200 mΩ (milliohms) (optional equipment)
Current leakage not more than 300 µA (microamps) (per UL 60606-1-1 (optional equipment)
Batteries sufficiently charged (optional scale system)
Display housing is intact and not damaged (optional scale system)
Load cells are intact and not damaged (optional scale system)
Foot support knee knob mechanism functions and can be secured in place (optional equipment)
Foot support leg knob mechanism functions and can be secured in place (optional equipment)
Foot support extends to the full extended position and stops in the correct position (optional equipment)
Foot support self-tapping screws (6) are secured and are not stripped (optional equipment)
Scale calibrated. Recalibrate if necessary (optional scale system)
Upright X-ray cassette holder is working as intended (Prime X option)
No damage to the fowler skin and foot skin (Prime X option)
No damage to the head and foot trays (Prime X option)
Bolt and nut through each fowler pivot is not loose (Prime X option)
Product serial number:
Completed by:
Date:
1-54

Preventive maintenance

1115-409-005 REV E
www.stryker.com