Replacing The Batteries For The Scale System Option - Non-Electric Litter; Charging The Scale System Battery Pack Option - Electric Litter Option Or Electric Litter Option With Chaperone Option - Not Available With Prime X Option; Arming Or Disarming The Chaperone Option - Stryker Prime Series Stretcher Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 147
Scale system option (Continued)
Do not touch the product while the scale system option is weighing or zeroing.
The patient must remain still while the system is weighing. If the patient is moving, the scale system will try for 20
seconds to get a stable weight or zero value before displaying the error message (
The error (
) appears in the display if there is a loose connection or if the scale is not operating. If you attempt to
operate the scale again and the error appears, contact Stryker Technical Support.
Note: To meet the scale accuracy claim, the patient surface must be in the flat position (Fowler backrest and gatch
down). The product cannot exceed 5 degrees of Trendelenburg/Reverse Trendelenburg.

Replacing the batteries for the scale system option - non-electric litter

Replace the batteries when the battery charge indicator displays one unit remaining. This prevents the scale system
option from shutting down due to drained batteries.
When the display flashes Lo batt, the batteries are at their lowest charge and the scale system option is disabled.
To replace the batteries:
1.
Use a Phillips screwdriver to remove the Phillips head screws that hold the battery compartment cover to the display
assembly.
2.
Remove all four AA batteries.
3.
Install four new AA batteries.
Always install the positive and negative poles as directed in the battery holder.
Only use Alkaline type (LR6) batteries.
Never mix old batteries with new batteries.
Always dispose of old batteries in accordance with local regulations.
4.
Use the screws removed in step 1 to reinstall the battery compartment cover on the display assembly.
Charging the scale system battery pack option - electric litter option or electric
litter option with Chaperone option - not available with Prime X option
The scale system with the electric litter option requires one 10.8V Li-ION battery pack (0058-135-000 without the
Chaperone option; 0058-134-000 with the Chaperone option). Recharge the battery pack when the battery charge
indicator displays one unit remaining. This prevents the scale system option from shutting down due to drained batteries.
To recharge the scale system Li-ION battery pack, plug the power cord into a grounded, hospital grade wall outlet. The
battery pack fully charges in approximately three hours.
Note: Always plug the product into a grounded, hospital grade wall outlet when not in use. This helps the battery to
maintain a sufficient charge and maximizes product performance while on battery power.

Arming or disarming the Chaperone option

When armed, the Chaperone option monitors the patient's position on the product.
To arm and disarm the Chaperone option:
1.
With the patient on the product, press Arm/Disarm.
Note: This arms Zone 1. Zone 1 illuminates amber when applied.
2.
To arm Zone 2, press Arm/Disarm twice within 3 seconds.
Note: This arms Zone 2. Zone 2 illuminates amber when applied.
www.stryker.com
Operation
1115-409-005 REV E
).
1-33