Rated maximum output power of
transmitter
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects, and people.
Guidance and manufacturer's declaration - electromagnetic immunity
The Optional Electric Lift or Electric Litter and Optional Scale System is intended for use in the electromagnetic
environment specified below. The customer or the user of the Optional Electric Lift or Electric Litter and Optional
Scale System should assure that it is used in such an environment.
Immunity Test
Electrostatic Discharge
(ESD)
IEC 61000-4-2
Electrostatic fast transient/
burst
IEC 61000-4-4*
Surge
IEC 61000-4-5*
1-56
EMC information
W
150 kHz to 80 MHz
d=1,2√P
0,12
0,38
1,2
3,8
12
IEC 6100 Test Level
±8 kV contact
±15 kV air
±2 kV for power supply
lines
±1 kV for input/output lines
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Separation distance according to frequency of transmitter
80 MHz to 800 MHz
d=1,2√P
0,12
0,38
1,2
3,8
12
Compliance Level
±8 kV contact
±15 kV air
±2 kV for power supply
lines
±1 kV for input/output lines
±1 kV line(s) to line(s)
±2 kV line(s) to earth
1115-409-005 REV E
m
800 MHz to 2.7 GHz
d=2,3√P
0,23
0,73
2,3
7,3
23
Electromagnetic
Environment Guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.