Support surface care (Continued)
WARNING (CONTINUED)
Always make sure that you wipe each product with clean water and thoroughly dry each product after any
application of chemical solutions. Some chemical solutions are corrosive in nature and may cause damage to the
product if you use them improperly. If you do not properly rinse and dry the product, a corrosive residue may be left
on the surface of the product that may cause premature degradation of critical components. Failure to follow these
instructions may void your warranty.
CAUTION
Do not allow liquid to seep into the zipper area or watershed cover barrier when washing the underside of the
support surface. Fluids allowed to come in contact with the zipper may leak into the support surface.
Do not iron, dry-clean, or tumble dry the support surface covers.
Do not power wash the support surface as this may damage the product.
Always completely dry the support surface covers before storing, adding linens, or placing a patient on the surface.
Drying the product helps to prevent the performance of the product from being impaired.
Do not over expose the covers to higher concentration chemical solutions as these may degrade the covers.
Do not use accelerated hydrogen peroxides or quaternaries that contain glycol ethers as they may damage the
cover and reduce the legibility of the graphics.
Failure to follow manufacturing instructions may also affect useful life of the support surface cover.
The support surface cover is resistant to the following chemical solutions:
Quaternaries (active ingredient - ammonium chloride)
Phenols (active ingredient - o-phenylphenol)
Chlorinated bleach solution (use 1 part bleach (5.25% sodium hypochlorite) to 10 parts of water which equals 4773
ppm of available chlorine (400 mL of a 5.25% bleach solution per 4000 mL water ))
70% isopropyl alcohol
Follow hospital protocol for support surface care in between patients to avoid the risk of cross-contamination and
infection.

Preventive maintenance

At a minimum, check all items listed during annual preventive maintenance for all Stryker Medical products. You may
need to perform preventive maintenance checks more frequently based on your level of product usage.
Remove product from service before performing preventive maintenance.
Note: Perform support surface care before inspection, if applicable.
Inspect the following items:
Zipper and covers (top and bottom) are free of tears, cracks, cuts, holes, or other openings
Internal components for signs of staining from fluid ingress or contamination by fully unzipping the covers
Labels for legibility, proper adherence, and integrity
Foam has not degraded or come apart
Fire barrier for rips or other visible signs of damage (If excessive wear of the fire barrier is observed,
replacement is recommended)
www.stryker.com
Operation
1704-009-005 REV C
English
EN
1-9