Herunterladen Inhalt Inhalt Diese Seite drucken

Stryker Prime Series Stretcher Bedienungsanleitung Seite 28

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 147
English
Scale button identification (Continued)
EN
Button
A
B
C
D
E
F
G
1-16
Introduction
Name
Power LED
Zone 1
Zone 2
Arm /Disarm
Weigh
lb/kg
Zero
1115-409-005 REV E
Function
Illuminates green when you
plug the power cord into a
wall outlet.
Illuminates amber when
you unplug the power cord
from the wall outlet.
Battery backup
functionality will turn on.
The patient can move
freely, but the alarm
sounds when the patient
attempts to leave the
product.
The patient can move with
limited movement. The
alarm sounds when the
patient approaches the
siderails or moves toward
the foot end.
Arms and disarms
Chaperone option
(stretcher exit)
Weighs the patient. The
display shows the patient's
weight for approximately
40 seconds before turning
off
Changes the displayed unit
of measure
Zeros the scale settings.
The display shuts off after
approximately 40 seconds.
Action
Push twice within 3
seconds to arm Zone 2.
Once armed or when
alerting, press once to
disarm.
Press and release the
Weigh button.
Push to toggle between
units of measure (lb or kg)
Push and hold the Zero
button for 2 seconds to
zero the scale system
before you put a patient on
the product. If the display
flashes HOLD, press and
hold the Zero button again
until the display reads REL.
Release the Zero button.
The display will flash 000.0
and then will display 000.0
(not flashing). The system is
not zeroed until the 000.0
stops flashing.
www.stryker.com

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Prime seriePrime x

Inhaltsverzeichnis