Basic Settings - LumaSense Technologies mpac IN 140/5-L Betriebsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

IMPAC pyrometers IN 140/5 · IN 140/5-H · IN 140/5-L
10.6

Basic settings

Under pyrometer parameters all preset values can
be displayed and modified if necessary.
The window pyrometer parameter contains all parameter
settings described in 9, Parameter descriptions / settings.
Choose the correct setting for your application, the actual
setting is displayed.
Notes:
"Basic range" displays the total range of the pyrometer
automatically and can not be changed. If the sub range is
changed the new values must be confirmed with "OK".
Under "Material" you have the possibility to store the
names of different measuring objects with their emissivity
values and to recall them from the list.
Choose whether the temperature should be displayed in
°C (Celsius) or °F (Fahrenheit).
If the pyrometer is equipped with a
laser targeting light the laser icon ( )
is visible. A click on the laser targeting
light icon turns the laser targeting light
on or off at this point. After approx. two minutes the laser targeting light is switched off automatically.
The open / save button enable to store and recall own pyrometer configurations.
„1 meas." shows the current measuring temperature in the pyrometer parameters window for
approx. 1 second.
A click on the "Test" icon opens a window that
allows the direct communication with the py-
rometer via the interface commands (see sec-
tion 14, Data format UPP
After entering an interface command (00 is the
adjusted address ex works, "ms" is the com-
mand "reading temperature value") and a click
on "Send" the following window is opened:
This window already shows the answer of the
1
pyrometer in
/
°. The actual temperature
10
reading is 325.7°C.
"Len" indicates the length of the answered data string, incl. Carriage Return (Chr(13)).
In the lower part of the window the connection with the preset baud rate can be checked. Here the command
was send 500 times with 19200 baud. It has taken 4.56 seconds without transmission errors.
If the pyrometer parameters window displayed, changing of settings on the pyrometer
Note:
is blocked.