IMPAC pyrometers IN 140/5 · IN 140/5-H · IN 140/5-L
14
Data format UPP
Via interface and a suitable communication software or via "Test" function of the InfraWin software (see 10.6
Test) commands can be exchanged directly with the pyrometer.
Basic settings
The data exchange occurs in ASCII format with the following transmission parameters:
The data format is:
8 data bits, 1 stop bit, even parity (8,1,e)
The device responds to the entry of a command with: output (e.g. the measuring value) + CR (Carriage Re-
turn, ASCII 13), to pure entry commands with "ok" + CR.
Every command starts with the 2-digit device address AA (e.g. "00"). This is followed by 2 small command
letters (e.g. "em" for level of emissivity
This is followed, if necessary for that command, by the ASCII parameter "X". If this parameter "X" is omitted,
then the device resets with the current parameter.
A „?" after the small command letters answers with the respective settings (only at setting commands, not at
enquiry commands).
Entry: "00em" + <CR>
Example:
The emissivity setting (
Answer: "0970" + <CR> means Emissivity = 0.97 or 97.0%
Description
Reading temperature
value:
Reading temperature
value repeated:
Emissivity:
Exposure time t
:
90
Clear times of the
maximum value
storage:
External clearing:
Analog output:
Reading basic tempera-
ture range:
Reading temperature
sub range:
Setting of temperature
sub range:
Address:
Baud rate:
Temp. display °C / °F
Wait time:
Internal temperature:
Max. internal
temperature:
®
(Universal Pyrometer Protocol)
ε
), finished with CR
ε
) of the device with the address 00 is returned
Command
Parameters
AAms
Output: XXXXX (dec., in 1/10 °C or °F)
AAmsXXX
XXX = 000...999 (XXX = number of measuring values)
AAemXXXX
XXXX = (0010 ... 1000‰) (decimal)
AAezX
X = 0 ... 6 (decimal)
AAlzX
X = 0 ... 9 (dez.)
AAlx
Clearing the max. value storage (only with lz = 7, external)
AAasX
X = 0 or 1
AAmb
Output: XXXXYYYY (hex 8-digit, °C or °F)
AAme
Output: XXXXYYYY (hex 8-digit, °C or °F)
AAm1XXXXYYYY
XXXX (hex 4-digit) beginning of temp. range (°C or °F)
YYYY (hex 4-digit) end of temp. range (°C or °F)
AAgaXX
XX = (00 ... 97)
AAbrX
X = 1...6 or 8 (dez.)
AAfhX
Output: X = 0: display in °C; X = 1: display in °F
AAtwXX
XX = 00 ... 99 (decimal, in bit time of the current baud rate)
AAgt
Output: XXX (dez. 000 ... 099°C or 032 ... 210°F
AAtm
Output: Output: XXX (dec. 000 ... 099°C or 032 ... 210°F)
(88880 = temperature overflow)
0 = intrinsic time constant of the device
1 = 0.01 s
3 = 0.25 s
2 = 0.05 s
4 = 1.00 s
0 = Maximum value storage off
1 = 0.01 s
4 = 1.00 s
2 = 0.05 s
5 = 5.00 s
3 = 0.25 s
6 = 25.00 s
0 = 0...20 mA; 1 = 4...20 mA
XXXX = beginning of temp. range
YYYY = end of temp. range
XXXX = beginning of temp. range
YYYY = end of temp. range
00 ... 97 = regular device addresses
99 = Global address with response
98 = Global address without response
(only setting commands!)
1 = 2400 baud
5 = 38400 baud
2 = 4800 baud
6 = 57600 baud
3 = 9600 baud
(7 is not available)
4 = 19200 baud
8 = 115200 baud
5 = 3.00 s
6 = 10.00 s
7 = external deletion
8 = automatically deletion
9 = Hold
23