Herunterladen Inhalt Inhalt Diese Seite drucken

Trouble Shooting - LumaSense Technologies mpac IN 140/5-L Betriebsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

IMPAC pyrometers IN 140/5 · IN 140/5-H · IN 140/5-L
13

Trouble shooting

Before sending the pyrometer for repair, try to find the error and to solve the problem with the help of the
following list.
Temperature indication too low
Incorrect alignment of the pyrometer to the object
⇒ New correct alignment to achieve the max. temperature signal (see 6)
Measuring object smaller than spot size
⇒ check measuring distance, smallest spot size is at nominal measuring distance (see 7)
Measuring object is not always in the measuring spot of the pyrometer
⇒ Use max. value storage (see 9.3).
Emissivity set too high
⇒ Set lower correct emissivity corresponding to the material (see 9.1)
Lens contaminated
⇒ Clean lens carefully (see 12.2)
Temperature indication too high
Emissivity set too low
⇒ Set lower correct emissivity corresponding to the material (see 9.1)
The measurement is influenced by reflections of hot machine parts
⇒ Use mechanical construction to avoid the influence of the interfering radiation
Measuring errors
Indicated temperature is decreasing during the use of the pyrometer, contamination of the lens
⇒ Clean lens. Recommendation: use of air purge (see 12.2, 5.1)
Indicated temperature is decreasing during the use of the pyrometer, although the air purge unit is used.
Probably compressed air is not clean or air failed
⇒ Clean the lens and use clean, dry and oil free compressed air
Air contamination in the sighting path between pyrometer and object
⇒ Change position of the pyrometer with a clean sighting path (if necessary use a ratio pyrometer)
HF-interferences
⇒ Correct the connection of the cable shield (see 4)
Instrument overheated
⇒ Use cooling jacket with air or water cooling (see 5.1)
Temperature Indication is fluctuating, probably caused by changing emissivity
⇒ Wrong pyrometer type, use of ratio pyrometer recommended
Laser targeting light
Laser targeting light fails
⇒ Instruments max. temperature of 50°C is exceeded. Use cooling jacket (see 5.1)
22

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Impac in 140/5-h'Impac in 140/5

Inhaltsverzeichnis