IMPAC pyrometers IN 140/5 · IN 140/5-H · IN 140/5-L
8

Instrument settings

The IN 140/5 is equipped with a wide range of settings for optimal adaption to the required measuring condi-
tion and for getting the correct measuring temperature (description of all available parameters see section 9,
Parameter descriptions / settings).
All instrument settings can be done directly at the instrument or via serial interface and software InfraWin
(see section 10, Settings via interface and software, user of an own communication software find all inter-
face commands on section 14, Data format UPP
8.1

Settings at the instrument

The LC-display as well as the push buttons for dis-
playing and setting of the parameters are found
inside the unit. The pyrometer is opened by 4 allen
screws. If unscrewed, the rear cover can be pulled
out along with the attached display and push but-
tons. The pullout is limited by the lengths of the
screws.
The backlight of the display is always powered in
either status, opened or closed pyrometer.
Please make sure that the py-
Note:
rometer is not contaminated
while open.
8.2

Key panel operation

With the PAR button all available parameters
1
PAR:
are displayed in the following description
(section 10). Pushing the button again
changes the display to the next parameter
and on the display a corresponding short
form is displayed (see section 9, in brackets
behind the parameter names).
With the arrow keys
2
:
settings can be displayed. Pushing the button
longer changes the settings in fast mode.
ESC / ENT: Pushing the ESC button changes the pyrometer to measuring mode. If a parameter is
3
changed with the arrow keys the indication of the ESC button changes to ENT. Pressing
the button again confirms the value into the pyrometer. Changing the parameters again by
pushing the PAR button doesn't confirm this value in the pyrometer. If no button is pressed
for 30 s the pyrometer changes to
changed value.
8.3

Selection of the serial interface

Opposite to the display
there is a switch to select
a serial interface RS232
or RS485. The LC display
shows as chosen either
RS232 or RS485.
®
)
and
all parameter
Interface
switch
Test
button
Max.
pull-out
Temperature dis-
play or parameter
short form
Push button
short description
the temperature indication without accepting the
Display of
the selected
interface
Screw connection
3 mm allen screw
Illuminated
Display
Setting keys
Rear cover
1
2
3
13