Information for service staff
10/2017
3.7
Disconnect the device from the power
supply and ensure that it cannot be
switched on again
DANGER!
Electricity throughout entire device!
If work is to be performed on the devices, disconnect
them from the power supply. This means that all live
cabling must be switched off.
wellsystem MEDICAL / MEDICAL_PLUS:
Disconnect the sunbed from the power supply when
performing maintenance work requiring opening of the
device.
wellsystem RELAX / RELAX _PLUS:
Before beginning work, pull out the mains plug or
disconnect all live lines.
Switching off the device alone is insufficient, as certain
points may still be live. Therefore, switch off all fuses
before performing any work and – if possible – remove
them.
Personal danger due to electric shock!
Severe accidents can occur if unintentionally switched on
again.
Immediately after disconnecting from the power supply,
ensure all switches or fuses used to disconnect the power
supply can not be switched on again.
–
Lock the fuse box with a padlock.
In case of circuit breakers which cannot be unscrewed, a strip of adhesive
tape may be applied over the actuation lever with the wording 'Do not switch,
danger!' (1 + 2).
A prohibition sign with the wording: 'Work in progress!'
'Place: ..............................'
'Sign may only be removed by: ............'
must always be securely displayed immediately.
DANGER!
Personal danger due to electric shock!
Prohibition signs must not be fixed to or come into
contact with live parts.
Installation instructions 13/68