Information for service staff
10/2017
3.2
Installation location
WARNING!
Salty air! Contaminated air!
Device damage due to corrosion of the housing and electri-
cal components.
–
Do not install the device at swimming pools.
–
Do not operate the device in locations with aggressive
environmental conditions (e.g. air containing chlorine)
Damage to device possible!
Where there are great differences in temperature between
the transport route and the installation site the device may
not be put into operation immediately after being set up.
–
Wait at least 2 hours before connecting the device to
the power supply.
Risk of overheating due to air flow problems!
Operating faults possible.
–
Do not change, obstruct or block the air inflow and
outflow to the device or make any unauthorized chang-
es to the device. The manufacturer shall bear no liabil-
ity for damage or injuries resulting from such actions.
–
Observe the ventilation system data.
–
Do not set up and operate the device on the transport
pallet!
Permissible Floor / Ceiling Loads
The massage unit has been designed to be installed on a flat and level sur-
face.
When installing the unit, you should always make sure that the floors and
ceilings in commercially used rooms are designed to support a load of
5000 N/m².
NOTE:
The load-bearing capacity of wooden beam ceilings must
be proven in individual cases.
If the actual maximum load exceeds this value, the operator must provide
separate documentary evidence in conformity with DIN EN 1991-1-1:2010-12
for the use of these rooms.
Example for ceiling load (based on normal room or cabin dimensions, 1
device including 1 user, 1 member of staff and small furnishings):
1 device up to 800 kg: Ceiling load approx. 2150 N/m².
Installation instructions 9/68