Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
61000-4-5
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
64/68
1009892-04
8.3.2
Table 202: Guidance and manufacturer's
declaration – electromagnetic immunity
The massage unit has been designed for use in the electromagnetic sur-
roundings listed below. The operator of this equipment must ensure that it is
only used in one of these surroundings.
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power supply lines ±2 kV for power supply lines
±1 kV differential mode
±2 kV common mode
3 A/m
<5 % U
(>95 % dip in U
)
T
T
for 0.5 cycle
40 % U
(60 % dip in U
)
T
T
for 5 cycles
70 % U
(30 % dip in U
)
T
T
for 25 cycles
<5 % U
(>95 % dip in U
)
T
T
for 5 s
Note: U
is the a.c. mains voltage prior to application of the test level.
T
Compliance level
±6 kV contact
±8 kV air
±1 kV differential mode
±2 kV common mode
3 A/m
<5 % U
(>95 % dip in U
)
T
T
for 0.5 cycle
40 % U
(60 % dip in U
)
T
T
for 5 cycles
70 % U
(30 % dip in U
)
T
T
for 25 cycles
<5 % U
(>95 % dip in U
)
T
T
for 5 s
Appendix
Electromagnetic
environment –
Guidance
10/2017