6 Commissioning
Distribution
Creating programmes - Event programme
Event prog.
Sample
Distribution
Switch
Service
Service
78
Set
Time:
Creating programmes
Bottle change after a specific time.
Switch 1,2
Number:
Distribution
Bottle change after a specific number of samples.
Ext. sig.:
Bottle change on external signal (connect signal input!).
Set
Sample:
Creating programmes
Sampling type in the event programme.
Event prog.
Distribution:
Distribution mode for the event programme.
Set
Time:
Creating programmes
Sampling acc. to time
Event prog.
Quantity:
Sample
Sampling acc. to volume (connect volume signal!).
mode
Ext. sig.:
Sampling on external signal (connect signal input!).
Flow:
Sampling acc. to flow (connect volume signal).
Once:
Single sample.
Set
Time:
Creating programmes
Bottle change after a specific time.
Event prog.
Number:
Distribution
Bottle change after a specific number of samples.
mode
Ext. sig.:
Bottle change on external signal (connect signal input!).
Set
No change
Creating programmes
Change:
Event prog.
Bottle change for event sample.
Distribution
1-x bottles:
Switch
Set
General:
Service
Software name, software version, software option, CPU number, pre-
set, view angle setting (contrast)
Run times:
Test: Service code necessary!!
Calibration: Calibration of the tap and the sample volume
Pump maintenance: Pump operation for changing the pump hose
– Run time - tube: operational time of the hose pump since the last
– Run time - limit: when the set operating time is reached, the warn-
reset;
ing 'CHANGE HOSE' appears on the display
Liquiport 2000 Ex
Time
-
Time
Time
No change
-
Endress+Hauser