6 Commissioning
Parameter
Info
Info
Basic settings
Name
Code
Date-time
Sampling
Inputs
Outputs
Memory interval
Interfaces
Programme selection
Number
Name
76
6.4.8
Description of device functions
The following table lists and describes all the setting parameters which you can read off and config-
ure when configuring the device.
Menu level
Set
Act. flow:
Info
Shows the current flow rate value of a connected flowmeter.
External stop:
Shows the number and duration of an external stop.
Event:
Shows the number and duration of event signals.
Set
Enter the required device name.
Basic settings
Name
Set
Enter the required user code.
Basic settings
!
Code
If you no longer know the user code, enter the customer code 6051 to
unlock the controller configuration.
Set
Date: Enter the current date.
Basic settings
Time: Enter the local time.
Date-time
Switch: Select the mode for switching between summer and winter
time.
Summer time: Date and value for switching to summer time.
Normal time: Date and value for switching to winter time.
Set
Time set.:
Basic settings
You can change the rinsing, suction, dosing and emptying times manu-
Sampling
ally.
Delay:
The sampling start (e.g. after an ext. signal) can be delayed by up to 99
seconds.
Rinse:
Number (0-3) of rinses before every sample.
Set
Function for digital input 1 and 2 and for setting the analog input.
Basic settings
Inputs
Set
Function for output
Basic settings
Outputs
Set
Time interval for internal recording of measured flow values. 0 sec = no
Basic settings
recording.
Memory interval
Set
Transmission rate and definition of the RS232 interface
Basic settings
Interfaces
Set
1: 1 main programme active
Programme selection
2: 1 main programme and 1 switching programme active
Number
1+Event:
1 main programme and one event programme active
2+Event:
1 main programme, 1 switching programme and 1 event programme
active
Set
For selecting the active sampling programme
Programme selection
Name
Setting options
Note!
Liquiport 2000 Ex
Factory setting
Liquiport 2000
0000
Customer code
6051
With automatic
switch between
summer and win-
ter time
Automatic
None
Power down
0000 sec
1
Programme 1-4;
1-U1; 1-E;
1-U1-E
Endress+Hauser