5 Operation
Tap calibration
64
5.6.2
Tap calibration
You must calibrate the tap if:
• The tap motor was replaced, or
• the error message <Tap calibration> appears in the display.
2.
3.
4.
A
5.
Calibration procedure:
1.Switch the device on
2.Under <Set -> Service - Calibration>, select the item
<distr. tab>.
3.When you select "Start", the tap turns and stops just
before the calibration position.
4.At the controller, keep selecting <1 step> until the
arrow on the front side of the tap is located exactly in the
notch in the middle of the distribution pan (point A).
5.In the picklist, select the menu option <save>.
6.Now the tab calibration is finished.
Liquiport 2000 Ex
Endress+Hauser