Herunterladen Diese Seite drucken

Endress+Hauser Liquiport 2000 Ex Bedienungsanleitung Seite 73

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Liquiport 2000 Ex
"
Endress+Hauser
Start-stop operation
Menu structure:
Fig. 23:
Start-stop operation menu structure
You can start the sampling programme either immediately by pressing the AUT key, or at a settable
time. You can specify the sampling programme stop as follows:
• Programme end: the unit stops sampling automatically when it has run through the set pro-
gramme.
Caution!
Danger of overfilling!
• No: the unit runs through the programme in a loop. Do not forget to empty the bottles.
• Time: the sampling programme stops at a settable time.
You can choose between continuous operation and operation at different intervals for the operating
mode.
• Day: operating time at two settable times per day.
• Week: operating time on three settable days per week.
• Interval: operating at specific time intervals.
Synchronisation
Menu structure:
Fig. 24:
Synchronisation menu structure
You can use synchronisation to assign specific filling times to specific bottles. In this way you could
fill bottle 1 between 00:00 and 02:00 and bottle 2 between 02:00 and 04:00 etc. For this, you have
the following options:
• AUT-button: the sampling and bottle change times are not synchronised.
• Synchronisation TIME: sampling starts with the first bottle. The change to the next bottles is syn-
chronised. E.g.: a time of 2:00 hours is set. For the synchronisation time of 00:00, the programme
is started at 05:20, for example, and bottle 1 is filled. The system then changes to bottle 2 at
06:00.
• Synchronisation TIME+BOT.: each bottle is allocated a specific filling time. Example: 00:00-
02:00: bottle 1; 02:00-04:00: bottle 2; 04:00-06:00: bottle 3, etc. For example if you start the
programme at 10:00, the unit will start filling bottle 6.