6.4
Device configuration
Note!
The figures show the menu structures with the display and picklist for the different configuration
parameters. Below the figures are tables containing the corresponding function descriptions for the
parameters.
6.4.1
Configuring the inputs
You can assign the inputs when configuring in the "Quick Setup" (see Section 6.3).
Digital inputs
Menu structure:
01:15
30.01.01
unit off
on
man
aut
on
..->
<-..
Set
Meld
Stat
Setup
Quick-Setup
Info
Basic Settings
Program Selection
Creating Programmes
Service
â
á
Esc
<-'
Fig. 15:
Digital input menu structure
Sample
Bottle
Quantity
Ext. stop
Event
Report
Switch
Basic Settings
Name
:
Port2000
Code
:
0000
Date-Time
Sampling
Inputs
Outputs
Multi-Sonde
mem. Interval
Interfaces
â
á
Esc
<-'
The input signal triggers a sample.
The input signal triggers a switch to the next empty bottle.
The input signal is a pulse signal from a volume meter; alternative to an analog
quantity signal (0/4...20 mA).
The input signal interrupts any programmes running; the programmes are contin-
ued when the signal disappears.
The input signal triggers an "event sample". For example, the input signal may be
switched when a limit value is infringed; a separate bottle can be filled in an event
sample.
A message appears in the display showing the date and time (e.g. function error of
flowmeter). You must acknowledge the message; the sample programme is not
interrupted.
The input signal triggers a change to the switching programme.
6 Commissioning
no
sampling
bottle
Inputs
quantity
ext. stop
Digital Input 1
event
Digital Input 2
report
Analog Input
switch
start aut
stop aut
aut 1x
â
á
Esc
<-'
67