Operation
98
Checking the oxygen measurement (sensor check system)
The oxygen value corresponds to a mV signal which is output by the oxygen sensor. Here, a certain
value for the oxygen is in a certain mV range. If the measured value is outside this range, the oxygen
sensor is most probably faulty and has to be replaced/serviced. It is possible to display the mV signal
output by the oxygen sensor with the sensor check system.
Menu structure:
Fig. 33:
Checking the oxygen measurement (sensor check system)
ASP station 2000 peristaltic