Operation
!
94
5.6.3
Calibrating the sensors of a connected multi-parameter probe
(optional)
Note!
For detailed information on calibrating the multi-parameter probe, please refer to the enclosed
Operating Instructions of the E+H multi-parameter probe CXS 70.
The multi-parameter probe is calibrated in the CALIBRATION menu level.
Menu structure:
Fig. 25:
Multi-parameter probe calibration
The menu structure for calibration is identical for the individual sensors and appears as follows:
Fig. 26:
Menu structure for sensor calibration
Act. value
Uncal
Cal. value
Calibrate
If the calibration is successful, the message 'CALIBR. SUCCES.' If the measured values are outside
specific limits, the message 'VALUE OUTSIDE' is displayed and the user is then prompted to
'ACCEPT CALIBRATION' or 'ABORT'. If a sensor cannot be calibrated, it may have to be replaced.
If sensors are calibrated too often, this can have a negative effect on the accuracy. For this reason,
it is possible to reset the sensors to the factory setting. In the sensor calibration menu in question,
Displays the current measured value of the sensor, the unit corresponds to the
setting made.
Calibration values are reset to the factory setting when you confirm with YES.
For entering the value of the calibration solution used.
The calibration is confirmed by selecting ACCEPT.
ASP station 2000 peristaltic