Commissioning
110
6.4.5
Creating a main programme
Menu structure:
Fig. 48:
Main programme menu structure
Sampling types
Menu structure:
Fig. 49:
Sampling types menu structure
You can take samples proportional to time, quantity, flow or controlled by events (see also "Sampling
types").
Time
Sampling is triggered after a settable time.
Quantity
Sampling is triggered after a specific measured flow rate.
!
For this function, a quantity signal must be connected to the analog input or one of the
digital inputs of the sampler. To do this, the digital input must be configured using the
QUANTITY function.
Ext. sig.
Sampling is triggered by an external signal.
!
One of the digital inputs must be assigned and configured using the SAMPLE function.
Flow
Sampling is triggered after a settable time. The volume sampled is proportional to a
measured flow rate.
!
A quantity signal must be connected to the analog input of the sampler for this function.
Programme name
The programme name can be changed as required with
the operating keys.
Note!
Note!
Note!
ASP station 2000 peristaltic