Commissioning
!
102
6.4
Device configuration
Note!
The figures show the menu structures with the display and picklist for the different configuration
parameters. Below the figures are tables containing the corresponding function descriptions for the
parameters.
6.4.1
Configuring the inputs
You can assign the inputs when configuring in the "Quick Setup" ( Chap. 6.3).
Digital inputs
Menu structure:
01:15
30.01.01
unit off
on
man
aut
on
..->
<-..
Set
Meld
Stat
Setup
Quick-Setup
Info
Basic Settings
Program Selection
Creating Programmes
Service
â
á
Esc
<-'
Fig. 38:
Digital input menu structure
Sampling
The input signal triggers a sample.
Bottle
The input signal triggers a switch to the next empty bottle.
Quantity
The input signal is a pulse signal from a volume meter; alternative to an analog quantity
signal (0/4...20 mA).
Ext. stop
The input signal interrupts any programs running; the programs will be continued when
the signal disappears.
Event
The input signal triggers an "event sample". For example, the input signal may be switched
when a limit value is infringed; a separate bottle can be filled in an event sample.
Report
A message appears in the display showing the date and time (e.g. function error of
flowmeter). You must acknowledge the message; the sample program is not interrupted.
Switch
The input signal triggers a change to the switching program.
Start Aut
The input signal restarts the program set in "Basic Settings-Program Selection". Any
program already active is reset.
Stop Aut
The input signal ends an active program.
Basic Settings
Name
:
Port2000
Code
:
0000
Date-Time
Sampling
Inputs
Outputs
Multi-Sonde
mem. Interval
Interfaces
â
á
Esc
<-'
ASP station 2000 peristaltic
no
sampling
bottle
Inputs
quantity
ext. stop
Digital Input 1
event
Digital Input 2
report
Analog Input
switch
start aut
stop aut
aut 1x
â
á
Esc
<-'