Commissioning
114
Sample mode
Menu structure:
Fig. 56:
Sample mode menu structure
The event programme contains the same sampling options (time, volume, ext. signal and flow) as
those of the main and switching programmes. In addition, you can select the function once. In this
function, the sampler takes a sample in the event programme once and then returns immediately to
the main programme.
Distribution
You can reserve separate bottles for the event programme. The following applies to bottle
distribution:
• The first bottle group of a distribution is reserved for the main programmes.
• The second bottle group is reserved for the switching programmes.
• The last bottle group is reserved for the event programme.
Menu structure:
Fig. 57:
Distribution menu structure
You can define the bottles for the event programme as follows:
No change
There is no bottle change when the system switches to the event programme.
Move
The next empty bottle is filled when the system changes to the event programme.
1-9 bottles
When the system switches to the event programme, 1-9 bottles are filled from the last
distribution bottle group. The number of bottles you can reserve for the event programme
is dependent on the total number of bottles (max. 9 bottles).
ASP station 2000 peristaltic
Endress+Hauser