Herunterladen Diese Seite drucken

Endress+Hauser ASP Station 2000 Peristaltik Betriebsanleitung Seite 115

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

ASP station 2000 peristaltic
Parameter
Info
Set 
Info
Info
Basic settings
Set 
Name
Basic settings 
Name
Set 
Code
Basic settings 
Code
Set 
Date-time
Basic settings 
Date-time
Set 
Sampling
Basic settings 
Sampling
Set 
Inputs
Basic settings 
Inputs
Set 
Outputs
Basic settings 
Outputs
Endress+Hauser
6.4.8
Description of device functions
The following table lists and describes all the setting parameters which you can read off and
configure when configuring the device.
Menu level
Ana. Input: Shows the current flow rate value of a connected
flowmeter.
External Stop: Shows the number and duration of an external stop
Event: shows the number and duration of event signals
Multi-probe: Shows the measured values of the probe if a multi-
parameter probe is connected (only with Connection for Multi-
parameter Probe option).
Enter the required device name.
Enter the required user code.
!
If you no longer know the user code, enter the customer code 6051
to unlock the controller configuration.
Date: enter the current date
Time: enter the local time
Switch: elect the mode for switching between summer and winter
time.
Summer time: date and value for switching to summer time.
Normal time: date and value for switching to summer time.
Time set.: You can change the purging, suction, dosing and emptying
times manually.
Delay: The start of sampling (e.g. after an external signal) can be
delayed by up to 99 seconds.
Flush: Number of flushes (0-3) before each sampling.
No Flow: With the "Counting" and "Off" parameters, the user can
specify whether the number of samples without flow should be
counted or not.
"
If the sample volume available is too low, the sampling is also aborted
due to "No Flow". However, the bottle has already been filled with
liquid. This is not taken into account when calculating the overfill
protection.
Man. Mode: Parameter specifies how the peristaltic pump should
react when the Man button is pressed.
"Sample": A sample is taken.
"Continuous": The pump is switched on and draws in the medium
until the Man button is actuated a second time.
Sample on ProgStart:
"Yes": A sample is taken each time the program is started or changed.
"No": No sample is taken each time the program is started or changed.
Function for digital input 1 and 2 and for setting the analog input.
Function for output
Setting options
Note!
Caution!
Commissioning
Factory setting
ASP2000 P
0000
Private code 6051
With automatic switch
between summer and
winter time
Automatic
00 sec
0
Counting
Sampling
No
No
No
115

Werbung

loading