Herunterladen Inhalt Inhalt Diese Seite drucken

Maintenance; Troubleshooting List - B. Braun Aesculap Acculan 4 GA320 Gebrauchsanweisung/Technische Beschreibung

Ladegerät
Vorschau ausblenden Andere Handbücher für Aesculap Acculan 4 GA320:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 12
8.

Maintenance

For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
9.

Troubleshooting list

Have defective products repaired by Aesculap Technical Service, see
Technical Service.
Fault
Charger not working
Defective indicator
Battery does not charge
Detection
"Power On" indicator not illumi-
nated
During the switch-on check, not all
indicator elements are illuminated
("Power On" indicator and indicator
fields for charging bays )
Battery inserted, charging progress
indicator not illuminated
Battery inserted, "Action request"
indicator illuminated
Cause
Mains power cord not plugged in
Power cable defective
Fuses blown
Charger defect
Charger defect
Charger contacts dirty
Battery defective
Charger defect
Excessive battery temperature
detected during charging
Battery defect
Charger defect
Remedy
Plug mains power cord into the
device connector 7 on the charger and
into the socket of the mains power sup-
ply.
Replace power cable.
Replace fuses, see Replacing fuses.
Replace product.
Replace product.
Clean the contacts, see Validated repro-
cessing procedure.
Repeat the charging process in another
charging bay. If the problem occurs
again, replace product.
The charging process does not start up
even with a different battery. Replace
product.
Remove the battery from the charging
bay, let it cool down, and restart charg-
ing. If the problem occurs again, replace
battery.
Repeat the charging process in another
charging bay. If the problem occurs
again, replace battery.
The charging process does not start up
even with a different battery. Replace
product.
en
7

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis