10. Replacing fuses

DANGER
Risk of death by electric shock!
Pull the mains plug before replacing the fuse sets.
Mandatory fuse set: 2x – T 1.25 AH, 250 V
Use a small screwdriver to release the clip on the fuse holder.
Pull out the fuse holder.
Replace both fuse sets.
Reinstall the fuse holder so that it audibly clicks into place.
Note
If the fuses burn out frequently, the device is faulty and should be repaired,
see Technical Service.

11. Technical Service

DANGER
Danger to life of patients and users in case of malfunctions and/or
failure of protective measures!
Do not modify the product.
CAUTION
Modifications to medical equipment may result in loss of guaran-
tee/warranty claims and any approvals granted.
Contact our national B. Braun/Aesculap representative for service
and repair.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Or in the US:
Aesculap Inc.
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042 USA
Aesculap Repair Hotline
Phone:
+1 (800) 214 -3392
Fax:
+1 (314) 895 -4420
Other service addresses can be obtained from the address indicated above.
8

12. Accessories/Spare parts

Art. no.
Name
TA020112
Fuse link T 1.25 AH, 250 V
TA014534
Instructions for use of Acculan 4 charger GA320 (A4 for
ring binder)
Power cable
Item no.
Approvals
TE780
Europe
TE730
Europe
TE734
Great Britain
TE735
USA, Canada, Japan

13. Technical data

13.1 Performance data, information about standards

Acculan 4 Charger GA320
Classification according to
Regulation (EU) 2017/745
Protection class according
to IEC/DIN EN 60601-1
Housing protection rating
according to
IEC/DIN EN 60529
For charging Aesculap bat-
tery types
Mains voltage ranges
(Current intake)
Frequency
Charging voltage/Output
voltage
Charging current/Output
current
Operational mode
Device protection
Weight without battery
Dimensions (L x W x H) with-
out battery
Dimensions (L x W x H) with
battery
Compliance with standards
EMC
CISPR11
Color
Length [m]
Black
1.5
Black
5.0
Black
5.0
Gray
3.5
I
I
IP21
GA346 and GA666, GA676
100 V~ to 120 V~ (0.8 A to 0.7 A) /
220 V~ to 240 V~ (0.5 A to 0.4 A)
50 Hz to 60 Hz
max. 15 V
max. 1.2 A ±0.3 A
Continuous operation
T 1.25 AH, 250 V
Form factor: 5 mm x 20 mm
approx. 2.3 kg
410 mm x 120 mm x 160 mm ±5 mm
410 mm x 252 mm x 160 mm ±5 mm
IEC/DIN EN 60601-1
IEC/EN 60601-1-2
Class A