QuantumX driver for CANape
7
Messages and information regarding error rectification
7.1 Error messages when opening a project
Error message
<Device name>: Error during import
settings from assistant!
<Device name>: Only one instance
allowed!
<Device name>: Could not start
QuantumX Assistant! Please install it!
7.2 Messages during a measurement
Message
<Device name>: Error start
measurement. Devices are
not synchronized!
<Device name>: Provided
buffer size too small, samples
lost!
A3325-1.0 en/de
Description
The driver cannot import the configuration file
(XML file) exported from the QuantumX
Assistant.
This error occurs when a CANape project is
opened where no configuration has so far
been implemented with the Assistant.
The QuantumX driver can only be integrated
(defined) once.
This error occurs when you try to define
another device that also uses the QuantumX
driver.
The QuantumX Assistant is not installed on
the PC or cannot be found.
Install the QuantumX Assistant again.
Explanation
A measurement cannot be started as long as a
QuantumX module is not synchronized.
Please note that the initial synchronization on an NTP
time server can take up to two hours. Synchronization
after a new switch on can take up to five minutes.
The storage space provided by CANape is not sufficient
and measured values have been lost. Data acquisition
for the measurement signal concerned will be terminated.
31