QuantumX driver for CANape
Enter the values for two points on the characteristic curve that you can
implement as a load on the sensor (Physical y1 and y2) and measure the
corresponding electrical values (Electrical x1 and x2). To do this, click on
The values for 0 N (no load) and 3000 N load were measured in the above
example. The entire measuring range of the sensor (Physical span) is
5000 N.
All changes to the channel settings are noted in the log. You can call up the
log via Show log top tight above the Ribbon.
5.7 Perform zero balance
In the Channels tab, mark the channels to be set to zero and then click on
Zero in the Zero group.
In order to delete a zero value again, click on Clear in the Zero group.
5.8 Setting the data rates and filters
Call up the Signals tab. In the default setting, just one signal per channel is
displayed (Options -> Hide all signals 2 in the Signals group), as CANape
can only process one signal per channel.
Double‐click in the relevant channel on the entry in the column Filter @ Data
rate to display the configuration dialog (Fig. 5.13).