87
Anhang
Recommended separation distances between portable and mobile RF communications equipment and
The UNIDRIVE
®
S III model 20 7010 20-1 is intended for use in an electromagnetic environment in which the RF disturbances are controlled. The customer
or user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the unit as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the
equation belonging to the relevant column, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note: These guidelines may not apply in all situations. The propagation of electromagnetic quantities is affected by absorptions and reflections of buildings,
objects and people.
Appendix
Table 206
the UNIDRIVE
®
S III model 20 7010 20-1
Separation distance d [m]
as a function of the transmitter frequency
150 kHz to < 80 MHz
80 MHz to < 800 MHz
D
0.12
0.38
1.2
3.8
12
Приложение
800 MHz to 2.5 GHz
D
F
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23