Electromagnetic Compatibility

Like other electrical medical equipment, SDC Classic™ requires special
precautions to ensure electromagnetic compatibility with other electrical medical
devices. To ensure electromagnetic compatibility (EMC), SDC Classic must
be installed and operated according to the EMC information provided in this
manual.
The SDC Classic has been designed and tested to comply with
IEC 60601-1-2:2001 requirements for EMC with other devices. It has not been
tested for immunity to electromagnetic disturbances.
Caution
Portable and mobile RF communications equipment may affect
the normal function of the SDC Classic.
Caution
Do not use cables or accessories other than those provided with
the SDC Classic, as this may result in increased electromagnetic
emissions or decreased immunity to such emissions.
Caution
If the SDC Classic is used adjacent to or stacked with other
equipment, observe and verify normal operation of the SDC
Classic in the configuration in which it will be used prior to using
it in a surgical procedure. Consult the tables below for guidance
in placing the SDC Classic™.