Setup and Interconnection

Stryker Endoscopy considers instructional training an integral part of the SDC
Classic™ system. Your Stryker Endoscopy sales representative will perform at
least one in-service at your convenience to help you set up your equipment and
instruct you and your staff on its operation and maintenance. Please contact
your local Stryker Endoscopy representative to schedule an in-service after your
equipment has arrived.
1.
Choose a location for the SDC Classic.
2.
Connect the AC power cable.
Warning
To avoid the risk of electric shock, this equipment must only be
connected to a supply mains with protective earth. To isolate
the device from the supply mains, use only hospital-grade
power cords.
3.
Connect peripheral devices according to the interconnection diagrams
on the following pages.
Warning
The SDC Classic is a non-sterile device intended for use outside
the patient environment. When setting up the device, select a
location that will ensure optimal ease of use while maintaining
separation between the device and the patient.
Do not connect the SDC Classic to extension cords or multiple
portable socket outlets (power strips) other than those built
into the Stryker cart.
Do not connect items to the SDC Classic that are not intended
for use with the system.
RF and other mobile communications equipment may affect
the normal function of the SDC Classic. When placing the SDC
Classic, follow the instructions located in the "Electromagnetic
Compatibility" section of this manual.
• Place the SDC Classic on a Stryker cart or other sturdy
platform.
• Provide at least four to six inches of space around the sides
and top of the SDC Classic to ensure proper ventilation.
13