Herunterladen Inhalt Inhalt Diese Seite drucken

AirSep VisionAire 5 Gebrauchsanweisung Für Patienten Seite 84

Vorschau ausblenden Andere Handbücher für VisionAire 5:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 19
VisionAire™
Recommended separation distances between portable and mobile RF communications equipment and the VisionAire
The VisionAire is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the VisionAire can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the VisionAire as recommended below, according to the
maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
Rated maximum
output power of
transmitter
from 150kHz to 80MHz
W
d= 1.2 x √P
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in Watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Guidance and manufacturer's declaration – electromagnetic emissions
The VisionAire is intended for use in the electromagnetic environment specified below. The customer or the user of the VisionAire
should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
®
AirSep
Corporation
MN137-1 rev. D
(m)
from 80MHz to 800MHz
from 800MHz to 2.5GHz
d= 1.2 x √P
0.12
0.38
1.2
3.8
12
Electromagnetic Environment guidance
The VisionAire uses RF energy only for its internal
function. Therefore, its RF emissions are very low
Group 1
and are not likely to cause any interference in nearby
electronic equipment.
The VisionAire is suitable for use in all
Class B
establishments, including domestic establishments
and those directly connected to the public low-
voltage power supply network that supplies buildings
used for domestic purposes.
Class A
Complies
EN
EN
7.0
Symbols are frequently used on equipment in preference to words with the
intention of lessening any possibility of misunderstanding caused by language
differences. Symbols can also permit easier comprehension of a concept within a
restricted space.
The following table is a list of symbols and definitions that may be used with
d= 2.3 x √P
AirSep's VisionAire Oxygen Concentrator. These symbols are referenced from
the appropriate International Electro-technical Commission (IEC) standards:
0.23
0.73
Symbol
2.3
7.3
23
EN-27
EN-28
Symbols/Abbreviations
Description
ON (power switch on)
No smoking
Type B equipment
Warning – Describes a
hazard or unsafe practice that
if not avoided can result in
severe bodily injury, death or
property damage
Caution – Describes a hazard
or unsafe practice that if not
avoided can result in minor
bodily injury or property
damage
Note – Provides information
important enough to
emphasize or repeat
Consult the accompanying
documents
Use no oil or grease
VisionAire™
Symbol
Description
OFF (power switch off)
Do not disassemble
Consult instructions for use
Class II Device
Complies with the 93/42/EEC
directive drawn up by the
approved organization No.
0459
Safety agency for CAN/CSA
C22.2 No. 601.1 M90 for
medical electrical equipment
Keep unit and accessories dry
Proper disposal of waste of
electrical and electronic
equipment required
®