Operating The Optional Wheel Lock(s)
1.
To activate the optional wheel lock(s), press fully down on the pedal (A) until it stops and is resting firmly against the
surface of the wheel.
2.
To release the optional wheel lock(s), depress the upper face of the pedal with your foot or lift up with your toe under
the pedal. The upper portion of the pedal will rest against the caster frame when the wheel lock is released.
WARNING
Never apply the optional wheel lock(s) while a patient is on the cot. Tipping could occur if the cot is moved while a wheel
lock is applied, resulting in injury to the patient or operator and/or damage to the cot.
Never leave a patient unattended on the cot or injury could result. Hold the cot securely while a patient is on the cot.
Never install or use a wheel lock on a cot with excessively worn wheels. Installing or using a wheel lock on a wheel with
less than a 6" diameter could compromise the holding ability of the wheel lock, possibly resulting in injury to the patient
or operator and/or damage to the cot or other equipment.
CAUTION
Wheel lock(s) are only intended to help prevent the cot from rolling while unattended. A wheel lock may not provide
sufficient resistance on all surfaces or under loads.