POWER PRO Troubleshooting Guide Table of Contents
Check for proper operation after each step. When the problem is fixed, return the cot to service. If assistance is needed
at any time during troubleshooting, please contact a service technician at (800) 327-0770 or (269) 324-6500.
Intended Function
Hold height position
Lift litter (with or without patient)
Lower litter (with or without patient,
powered)
Extend base when unloading
(powered)
Retracts base when loading
(powered)
Extend base when unloading
(manual)
Retract base when loading
(manual)
Lower litter (with patient, manual)
Lower litter (without patient,
manual)
www.stryker.com
Troubleshooting Guide
Problem
Cot drifts down
Leaking hydraulic fluid
Does not lift (patient weight of 500 lbs (227 kg) or less)
Lifts slowly
Does not lift smoothly
Does not lift to set load height
Excessive motor noise
Does not lower
Lowers too quickly
Does not lower smoothly
Cot is noisy while lowering
Base does not extend
Base extends slowly
Base extends too far
Base doesn't extend far enough
Base doesn't extend smoothly
Base compresses significantly after removing load wheels
from ambulance
Excessive motor noise
Base doesn't retract
Base retracts slowly
Base doesn't retract smoothly
Base doesn't stay retracted (drifts down)
Excessive motor noise
Does not allow extend
Extends too slowly
Does not allow retract
Does not lower
Does not lower smoothly
Does not lower
Does not lower smoothly
6500-009-005 REV A
See Page,
Section
67, 3
67, 4
67, 1
67, 1
68, 5
68, 6
68, 7
68, 8
68, 9
68, 9
68, 5
67, 1
67, 3
68, 10
68, 10
68, 5
68, 5 and-
68, 11
68, 7
69, 12
69, 12
69, 13
68, 11
68, 7
69, 14
69, 14
69, 15
69, 16
69, 16
69, 17
69, 17
6