Note A charged AHTO Irrigation Pump will run on batteries for
approximately 30 cases at the high flow rate. When the batteries are
low, the bottom light on the battery level display will flash red,
indicating enough battery life for one typical case and the need for
recharging.
Note To run the console using the battery pack, wind and store the cord in
the cord holder.

Operating the Unit

1. Remove the tube set (P/N 250-070-600 or
250-070-620) from its pouch.
2. Spike the Irrigation Bag. Ensure that the
distance between the console and the bag
gives the tubing some slack.
3. Connect suction adapter to wall suction.
4. Insert cassette. Ensure both tabs on the
cassette are properly latched onto the
pump. (The latches will sound 2 "clicks").
Note The system powers on and off
automatically when the cassette is
inserted and removed from the
pump.
5. Select the desired flow rate.
L = 2L/min.
M = 3L/min.
H = 4L/min.
6. Attach the appropriate irrigation tip.
WARNING Ensure the irrigation tip is
Note Follow proper hospital protocols for
use of electrosurgical equipment and probes.
7. To prime the system, ensure the handpiece is at least 18" below the
Irrigation Bag. (24" and greater is recommended).
free from damage or rough
surfaces prior to use.