Rotor-Gene AssayManager Core Application Handbuch – Oktober 2012
Fehler-
kennung Fehlertext
510014
The communication with
the cycler was lost. Switch
off the cycler, switch it back
on again. If error persists,
contact QIAGEN Technical
Services.
510018
The rotor has stalled or
stopped.
510019
The machine's source
motor jammed. Contact
QIAGEN Technical
Services.
510020
The cycler is taking too
long to reach temperature.
This can affect the assay
performance.
510021
The temperature
measurement thermistor
has gone open circuit.
Contact QIAGEN Technical
Services.
510022
Undefined error. Switch off
the cycler, switch it back on
again, and restart the
application. If the error
persists, contact QIAGEN
Technical Services.
Rotor-Gene AssayManager Core Application Handbuch 10/2012
Die Kommunikationsverbin-
dung mit dem Thermocycler
besteht nicht mehr. Schalten
Sie den Thermocycler aus
und anschließend wieder ein.
Falls der Fehler weiterhin
besteht, kontaktieren Sie den
Technischen Service von
QIAGEN.
Der Rotor blieb stehen oder
wurde angehalten.
Der Motor der Anregungs-
Strahlenquelle des Geräts ist
blockiert. Kontaktieren Sie
den Technischen Service von
QIAGEN.
Der Thermocycler braucht zu
lange, um die Temperatur zu
erreichen. Dadurch kann die
Assay-Leistungsfähigkeit
beeinträchtigt sein.
Der Stromkreis des Tempe-
ratur-Messthermistors ist
offen. Kontaktieren Sie den
Technischen Service von
QIAGEN.
Nicht definierter Fehler.
Schalten Sie den Thermo-
cycler aus und dann wieder
ein, und starten Sie die
Applikation neu. Falls der
Fehler weiterhin besteht,
kontaktieren Sie den Techni-
schen Service von QIAGEN.