Rotor-Gene AssayManager Core Application Handbuch – Oktober 2012
Fehler-
kennung Fehlertext
430001
The required channel {0}
on the selected cycler could
not be found. The run
cannot be started.
430012
The run could not be
started on the cycler with
the serial number {0}.
Make sure the lid is closed.
430020
Persisting the experiment
failed. See error log for
details.
430021
Unknown error during run
profile execution.
430023
The merged contains a
wrong acquisition type:
{0}. Expected {1}. The run
cannot be started.
430024
Within one cycle, the
runProfileEntryIndex must
not change. The run cannot
be started.
430030
The run was stopped. For
more information see
experiment error log.
1-380
Rotor-Gene AssayManager Core Application Handbuch 10/2012
Der erforderliche Kanal {0}
des ausgewählten Thermo-
cyclers konnte nicht gefunden
werden. Der Lauf konnte
nicht gestartet werden.
Der Lauf konnte auf dem
Thermocycler mit der Serien-
nummer {0} nicht gestartet
werden. Stellen Sie sicher,
dass der Deckel geschlossen
ist.
Fortführen des Experiments
fehlgeschlagen. Weitere
Einzelheiten siehe Fehler-
Logdatei.
Unbekannter Fehler während
der Ausführung des Lauf-
Profils.
Die verbundene Zelle enthält
einen falschen Erfassungstyp:
{0}. Erwartet wurde {1}. Der
Lauf konnte nicht gestartet
werden.
Während eines Zyklus darf
sich runProfileEntryIndex nicht
ändern. Der Lauf konnte
nicht gestartet werden.
Der Lauf wurde angehalten.
Weitere Informationen finden
Sie in der Fehler-Logdatei
zum Experiment.