Rotor-Gene AssayManager Core Application Handbuch – Oktober 2012
Fehler-
kennung Fehlertext
110044
The run template of the
current assay profile does
not match the run settings
of the .rex-file.
110048
One or more steps in the
assay profile editor are
invalid. Correct these
invalid steps to start the
assay profile tester.
110049
The .rex file at {0} cannot
be accessed. Check if the
.rex file path is correct.
110055
The color channels of the
run profile of the current
assay profile do not match
the .rex file color channel
{0}. Select another assay
profile or .rex file.
110056
The samples of an assay
must be arranged without
gaps. At .rex file tube
position {0} on page {1} a
sample is positioned after a
gap.
Rotor-Gene AssayManager Core Application Handbuch 10/2012
Die Lauf-Vorlage des aktu-
ellen Assay-Profils stimmt
nicht mit den Laufeinstel-
lungen der .rex-Datei
überein.
Ein oder mehrere Schritte im
Assay-Profil-Editor sind un-
gültig. Korrigieren Sie diese
ungültigen Schritte, um den
Assay-Profil-Tester zu starten.
Auf die .rex-Datei in {0}
kann nicht zugegriffen
werden. Überprüfen Sie, ob
der Pfad der .rex-Datei
korrekt ist.
Die Farbkanäle des Lauf-
Profils des aktuellen Assay-
Profils stimmt nicht mit dem
Farbkanal {0} in der .rex-
Datei überein. Wählen Sie
ein anderes Assay-Profil oder
eine andere .rex-Datei.
Die Proben eines Assays
müssen lückenlos angeordnet
werden. Bei Röhrchenposition
{0} auf Seite {1} in der .rex-
Datei ist eine Probe nach
einer Lücke angeordnet.